Optical Complex Spectrum Analyzer – Characterization of High Repetition Rate Optical Frequency Combs

October 2026

ABSTRACT

The ultra-low noise and GHz repetition rate of the Mehnir-1550 Optical Frequency Comb (OFC) series are important features for many applications such as precision spectroscopy. However, the individual comb line cannot be resolved with grating based Optical Spectrum Analyzers that offers spectral resolution of up to 10 pm.

The APEX Technologies High resolution Optical Spectrum Analyzer (OSA)/Optical Complex Spectrum Analyzer (OCSA) are perfect tools to verify and evaluate some important specifications of OFC sources. This equipment is based on interferometric method and is able to measure spectrums with the highest resolution in the market (5 MHz/ 40 fm), high wavelength accuracy (+/- 2 pm), for any wavelength range between 700 to 1700 nm. It has also the added benefit of measuring phase as a function of frequency. The phase and intensity informations can then be used to calculate phase, chirp, alpha parameter or pulse shape as a function of time.

In this application note, we present the characterizaton of the MENHIR-1550 laser with 400 MHz comb line spacing using the OCSA-AP6 from APEX TECHNOLOGIES.

APEX Technologies company logo in RGB colors

APEX Technologies – Optical Complex Spectrum Analyzer

  • Highest spectral resolution in the market
  • No Baud rate limitation
  • No modulation format limitation

Menhir Photonics – Optical Frequency Comb

  • Excellent passive repetition rate stability
  • Compact-all-in-one design

Figure 1. a) APEX Technologies High resolution Optical Complex Spectrum Analyzer for the C+L band (model OCSA-AP6-6). b) Mehnir-1550 Optical Frequency comb with 400 MHz comb line spacing. The APEX Technologies OCSA-AP6 measures the Intensity, phase and pulse width of the MENHIR-1550 comb laser.

HIGH RESOLUTION OPTICAL SPECTRAL MEASUREMENT

Thanks to our state-of-the-art technology, APEX Technologies equipment can measure not only the OFC envelope, but also the details. The OFC follows the sech2 fitting and the frequency spacing between the lines corresponds to the 400 MHz specified in the OFC datasheet.

Figure 2 - The ultra-low noise and GHz repetition rate of the Mehnir-1550 Optical Frequency Comb (OFC) series are important features for many applications such as precision spectroscopy. However, the individual comb line cannot be resolved with the grating based Optical Spectrum Analyzers that offers spectral resolution of up to 10 pm. The heterodyne principle used for spectrum characterization induce a particular peak shape with a ‘hole’ in the center (M-shape) because the heterodyne signal is detected with balanced photodiode to supress the DC noise.

PULSE WIDTH AND RELATIVE SPECTRAL PHASE MEASUREMENTS

The principle of OCSA is to measure the complex amplitude of every mode, represented by power and phase values in spectral domain. In addition to the spectral phase charactezition, this enables the calculation of the pulse width in the time domain by Fast Fourier Transform.

 

Figure 3. a) Measuring pulse width: we measured a pulse width of 600 fs, compared with the 200 fs specified by the manufacturer. This pulse stretching is explained by the use of a fiber and an optical attenuator. b) Measuring relative phase versus frequency: the relative phase of the Menhir-1550 comb is clearly linear.